●Compact design, save space in O.R. or Laboratory Room
●Built-in 5.5″ LCD Anaesthesia ventilator
●New compact CO2 absorber (autoclavable. Patent)
●High precise big volume vaporizer
●Optional: Bain circuit interface, Continuous FIO2 monitoring
●Optional: Battery back up for all system not less than 60 mins
Technical Specification
Anaesthesia main unit:
■High resolution 5.5” LCD screen
■Close, semi-closed, semi-open system
■Application: Adult, Pediatric
■Low gas alarming: Low O2 gas pressure, it will give alarming more than 7 seconds.
■High-pressure Limitation: Whenever airway pressure exceeds the pre-set High pressure, the ventilator offers protection automatically, to assure airway pressure can not increase anymore which supplies safety to patients.
■Low-pressure protection: Once O2 pressure less than 0.1MPa, the N2O gas supply was stopped automatically.
■Power failure alaming, power extendable function
■02flush: 25 ~ 75 L/min
■Gas Pressure: O2: 0.3 ~ 0.5 MPa .
N2O: 0.3~0.5 MPa
■Power: 220V+10%。 50Hz+1%;
Flow meter:
■O2, N2O – single tube each (0.1 ~ 10 LUmin)
■O2/N2O inter-lock system, to keep O2 concentration no less than 25%
Anaesthesia Ventilator:
■Adult bellow 1 set, Optional 1 set Pediatric bellow ;
■Pneumatic driven electric control
■Monitor Parameters: Tidal volume, Respiratory frequency, I: E rate,
Peak airway pressure, airway pressure wave (P-T)
■Working mode: Manual, Auto (IPPV、SIGH、 SIMV/FI2、 SIPPV)
■Tidal volume: 0, 50 ~ 1500 mL
■Respiratory frequency:.4 ~ 60 bpm
■I:E: 1:1~ 1:4 (adjustable)
■Synchronizer Tigger Pressure:-3 ~ 0KPa
■ High-pressure lmitation:2~ 6 KPa
■ Low-pressure limitation: 0.6 ~ 2 KPa
Respiratory Circuit:
■Consolidated compact Co2 absorber (autoclavable) with its own patent.
ZF-3000 Vaporizer:
■One precise vaporizer (Enflurane, Halothane, Isoflurane, Sevoflurane)
■Adjustable range: 0.5~5%VOL
■Automatic compensation function – flow, pressure, temperature
■Sel-lock function
■Volume: 280 ml