Features
· Pneumatically driven electrically controlled ventilator, can be used for emergency clinical resuscitation and respiratory therapy
· Display: High visibility, wide angle LED alphanumeric data display, equipped with tactile analog parameter controls
· Equipped with high precession infrared turbine pressure and flow sensors
· Power failure support functions: Automatically converts to standby mode during AC power outages
· Nonhazardous environmentally friendly frame, precise construction, elegant design, equipped with automated self-checking function during startup and an automatic self-calibrating sensor
· Alarms: Audible and/or visual alarms for tidal volume, airway pressure, power failure oxygen failure and other alarms.
· 9 Level adjustable humidifier
Specifications
Model | PA 500 |
Display | LED |
Ventilation Mode | A/C, IPPV, SIPPV, IMV, SIMV, MANUAL |
Minute Volume | Greater than or equal to 18L |
Back-up Power Supply | At least 4 hours |
Tidal Volume | Adjustable range: 50-1500ml, Display range: 0-2000ml |
Output Oxygen Concentration | <50% |
Respiratory Rate | 6-60bpm |
Pressure Limit | 1-6kPa |
Inspiratory/Expiratory (I:E) Ratio | 1:1.5-1:3.0 |
PEEP | 1-10cmH2O |
SIMV Rate | 1-12bpm |
Inspiratory Pressure Trigger | .-4-10cmH2O |
Maximum Safe Operating Pressure | ≤6.0 kPa |
SIGH | 1 sigh breath in every 80-120 breaths |
Airway Pressure Alarm | Upper Limit: 0.9~5.4kPa, Lower Limit: 0.5kPa |
Gas Supply Requirement | 280~600 kPa medical grade oxygen |