Common Features
Operating Modes: Pneumatically driven & electronically controlled, closed, semi-closed and semi-open
Pneumatically driven electrically controlled ventilator, can be used for emergency clinical resuscitation and respiratory therapy
Display: Built-in 10.4 inch color TFT fully adjustable display, able-display Pressure-Time, Frequency-Time, Pressure-Volume, Frequency-Volume and any two waveforms simultaneously on screen
Trigger Types: Pressure trigger, Flow trigger
Sigh: Ability to insert 1-8 sigh breaths in every 100 breaths, ventilation should be no less than 1.5 the tidal volume
Features built-in dual electronic flow sensors and dual respiratory modules
Power Failure Support Functions: Automatically converts-standby mode during AC power outages
Monitor Parameters: tidal volume, respiration rate, airway pressure, I:E ratio, inspiratory & expiratory status, inspiratory trigger and others
Equipped with infrared turbine flow sensor; precisely measures gas flows, stable and reliable
9 Level adjustable humidifier
Nonhazardous environmentally friendly frame, precise construction, elegant design, equipped with automated self-checking function during startup and an automatic self-calibrating sensor
Alarms: Audible and/or visual alarms for ventilation volume, tidal volume, oxygen concentration, asphyxia, airway pressure, intubation disconnection, low oxygen, power failure and other alarms, also features a alarm history query function
Advanced Model Exclusive Features
Equipped with professional medical grade air compressor tailored for long clinical times and low noise requirements
Air Compressor
Air Compressor: 220V AC±10%, 50±1Hz, 700VA, also features good protective grounding against electrical activity
Output Gas Pressure: 0.4MPa±20%
Continuous output gas flow greater than or equal to 20L/min
Instantaneous output gas flow great than equal to 60L/min
Specifications
Model | 900B (Standard Model) | 900B (Advanced Model) |
Display | 10.4 inch high visibility color TFT display | |
Ventilation Mode | A/C, SIPPV, IPPV, SIMV, PSIMV, PSV, CPAP, PCV, VCV, SPONT MANUAL | |
Minute Volume | Greater than or equal to 18L | |
Back-up Power Supply | At least 4 hours | |
Tidal Volume | Adjustable range: 20-1500ml, Display range: 0-2000ml | |
Respiratory Rate | 1-99bpm | |
Inspiratory/expiratory (I:E) ratio | 4:1-1:4 | |
PEEP | 0-10cmH2O | |
SIMV Rate | 1-20bpm | |
Inspiratory Pressure Trigger | .-10-10cmH2O | |
Trigger Sensitivity (flow) | 1-10L/m | |
Inspiratory Plateau | Adjustable from 0-50% of inspiratory time | |
Adjustable Oxygen Concentration | 45-100% | 21-100% |
Gas Supply Requirement | 280~600 kPa medical grade oxygen and compressed air supply | |
Pressure Limit | 1.0kPa~6.0kPa | |
Maximum Safe Operating Pressure | ≤6.0kPa | |
Tidal Volume Alarm | Upper Limit: 10~2000ml, Lower Limit: 0~1800ml | |
Airway Pressure Alarm | Upper Limit: 0.1~6kPa, Lower Limit: 0~5kPa | |
Oxygen Concentration Alarm | Upper Limit: 22~100%, Lower Limit: 21~80% |