Features
3.5 inch LCD screen for clear vision within 5 meters
Special designed extension line grip
One-button night mode design for patient’s more comfortable rest at night
Compatible with different brands of 5,10,20/30,50/60 ml syringes
User friendly numeric setting key
More stronger syringe fixing slice & clamp
Free stacking of single syringe pumps for easy space stations in ICU, NICU, OT etc
Display features
1 Battery display
2 Time left display
3 Drug library code display for clinical & customized drug use
4 Body-Weight mode to satisfy different drug dosage in clinical use
5 Rate,Time-Rate,Volume-Rate,Time-Volume to satisfy different clinical choices
Backside features
1 RS232
2 DC for ambulance or different condition power supply
3 Liquid drop cover to protect liquid into AC power supply socket
WE HAVE MORE
+ Module Design for easier local service in any countries
+ Double CPU to ensure higher reliability
+ General graphic identification
+ Manufacturing according to Chinese YYT0287 & European ISO13485
+ Rosh & CE qualified
Model No.
|
SPA112
|
SPA122
|
||
Channel
|
Single
|
Double
|
||
Stackable
|
YES
|
NO
|
||
Syringe Size
|
5.10,20.30.50/60ml
|
|||
Infusion Modes
|
Rate Mode, Rate-Time, Rate-VTBI, Time-VTBI, Body Weight
|
|||
Safety
|
Double CPU to make sure higher standard infusion
|
|||
Drug Library
|
20 drug list with drug code display
|
|||
Accuracy
|
±2%
|
|||
Time Preset
|
00:01-99: 59 (Hour: Min)
|
|||
Volume Range
|
0~9999.9ml
|
|||
Flow Rate Step by
|
0.1 ml/h when Rate <100ml/h. 1ml/h when Rate ≥ 100ml/h
|
|||
Flow Rate Range
|
5ml Syringe 0.1 ml/h-150ml/h
|
|||
10ml Syringe 0.1ml/h-300ml/h
|
||||
20ml Syringe 0.1ml/h-600ml/h
|
||||
30ml Syringe 0.1ml/h-900ml/h
|
||||
50/60ml Syringe 0.1ml/h-1200ml/h
|
||||
Purge/Bolus
Rate/Top Flow Rate
|
5ml Syringe 150ml/h; 10ml Syringe 300ml/h ; 20ml Syringe 600ml/h;
30ml Syringe 900ml/h ; 50/60ml Syringe 1200ml/h
|
|||
Audible & Visible Alarms
|
Auto Seft-Test, Syringe Dislocation, Occlusion, Near End, Syringe Empty, VTBI Completion, Low Battery, Battery Exausted, Motor Malfunction, Incorrect Syringe Specification, Circuit Malfunction, Master CPU Malfunction, Monitoring CPU Malfunction, Parameters Over Limitation, A.C Disconnection, A.C connection
|
|||
KVO
|
0.1-5.0ml/h Adjustable
|
|||
Occlusion Pressure
|
High: 40 KPa±20KPa
|
|||
Middle: 60 KPa±20KPa
|
||||
Low: 100KPa±20KPa
|
||||
Max Infusion Pressure
|
120KPa
|
|||
Battery
|
≥ 8 hours
|
≥ 4 hours
|
||
Lithium battery,11.1/2000mAh
|
||||
Power Consumption
|
30VA
|
45VA
|
||
Power Supply
|
AC: 100-240V, 50Hz/60Hz
|
|||
Classification
|
Class l, type CF, IPX4
|
|||
Dimension
|
26×21.5x11cm
|
32×21.5x20cm
|
||
Weight
|
2kg
|
3kg
|
||
Optional Function
|
Ambulance DC: 12V
|